Evaluation of Reject Analysis of Chest Radiographs in Diagnostic Radiology
Keywords:
Reject, analysis, Chest, Positioning errors, ArtifactsAbstract
Introduction: Reject analysis is an important quality indicator and is required for reducing radiation exposure to the patient and increased cost-effectiveness.
Aim: The study aims to analyze reject rates and causes in direct chest radiographs to assess the radiographic quality.
Materials and Methods: A total of 150 chest radiographs (CXR) had been retrospectively analyzed at King Fahad Hospital from January to March 2016.
Results: The rejection rate was 14.7%. Positioning errors and artifacts were the most significant factors for reject analysis in chest radiography. Positioning errors are the most common cause of rejection (16%), and artifacts are the second reason (11.33%). In consequence to positioning error, the scapular shadow was partially located at the lung field in 38% of the images.
Conclusion: chest radiographs are the most common x-ray examination requested. Positioning errors are still challenging, which affect the quality of the image. An overall rejection rate of 14.7% indicates a need for continuous practice in the Radiology department to improve the performance.
Publication and Publisher Transfer Disclaimer: This article was originally published by WisdomGale Publishing, having its registered address at 14 Rue de Grand-Bigard, 1082 Brussels, Belgium. Following the formal transfer of publication rights and the subsequent change of publisher, the article was migrated to the publishing platform of International Medical Publishing Group (IMPG), India. The original publication details, authorship, and scholarly content remain unchanged. All subsequent publication management and archival responsibilities are administered by IMPG, India, in accordance with the terms and conditions of the publisher transfer agreement.